Special Operation Preference setup » History » Version 5

Amber Herold, 06/22/2010 10:20 AM

1 1 Amber Herold
h1. Special Operation Preference setup
2
3
4
5
h2.  Taking final exposures at very high magnification
6
7
8
9 5 Amber Herold
Hysteresis of image shift at very high mags such as 200 kx can be reduced by setting the following configuration
10 1 Amber Herold
11
12
*  Presets Manager/Settings>
13
14
15
Cycle Magnification Only = no
16
17
18 3 Alex Kadokura
*  If [[Queuing Example 1 - Exposure Targeting|queuing is used at exposure targeting]],
19 1 Amber Herold
20
21
Exposure Targeting/Settings>
22
23
24
Declare drift when queue submitted = yes
25
26
27
28
29
30
31
32 5 Amber Herold
h2.  Avoiding drift correction that uses LM sq preset
33 1 Amber Herold
34
35 4 Alex Kadokura
36 5 Amber Herold
If the grid is flat and the offset of targets is small enough that it can be corrected in the hole image, we recommend using the [[Queuing Example 1 - Exposure Targeting|Queuing option at Exposure Targeting]]. The sq preset is still used after Stage Z adjustment to correct the hole targets but will not be used when queued exposure targets are processed.
37 1 Amber Herold
38
39 5 Amber Herold
If the offset of targets is minimal even after stage Z adjustment, set the following to avoid using ancestor images obtained in LM for drift management purpose, queuing or not.
40 1 Amber Herold
41
42
*  Target Adjustment/Settings>
43
44
45 5 Amber Herold
Minimum Magnification =Lowest Magnification you want to allow the node to use for determining the new target.
46 1 Amber Herold
47
48
49
50
h2.  Minimize the use of image shift for the final exposure (Approach 1)
51
52
53
54 5 Amber Herold
Large image shift can cause sufficient beam tilt that creates problems such as bad autofocus, beam shift, and loss of resolution (but probably only if better than 4 angstrom is needed). Image shift is used in Leginon for targeting exposure because it is much more accurate than a single movement by the specimen goniometer, even after it is modeled for mechanical periodicity.
55 1 Amber Herold
56
57 5 Amber Herold
The accuracy of the stage movement is lower when it moves a long distance. Therefore, using iterative stage movement can improve the targeting, and therefore reduce the amount of image shift required for the final exposure. To use this feature, change the following preference in Hole node (or Subsquare node in MSI-Raster).
58 1 Amber Herold
59
60 5 Amber Herold
It is important to know that in order to check whether the precision is reached, image is taken at the preset at which the target is selected on (in this case, sq preset), so the dose should be kept at minimal. On our microscope 0.2 micron precision can be obtained within 2 to 3 moves.
61 1 Amber Herold
62
63
Hole/Settings/Image Acquisition>
64
65
66
* Mover=navigator
67
68
69 5 Amber Herold
* Navigator Target Tolerance = 2 e-7 m (or whatever tolerance you like): This sets the goal for multiple movement
70 1 Amber Herold
71
72 5 Amber Herold
* Navigator Acceptable Tolerance = 1 e-6 m (or whatever tolerance you like): If further movement causes an increase rather than a decrease of targeting accuracy, this value determines whether the target is aborted or not.
73 1 Amber Herold
74
75
* Final Image Shift = No
76
77
78
79
80
Navigation/Settings/Error Checking and Correction>
81
82
83 5 Amber Herold
*  Preset cycle after each move = yes if your sq preset is in LM mode and you don't mind wait a little longer; = no if you don't have hysteresis problem staying in sq preset for the unknown amount of time during the iterative move.
84 1 Amber Herold
85
86
87
88
89
90
91
h2.  Minimize the use of image shift for the final exposure (Approach 2)
92
93
94
95 5 Amber Herold
A different approach is being developed to minimize image shift while maintaining final targeting accuracy-A combination of stage movement and image shift. This approach can only be used in limited cases such as targets selected for the tomography node or exposure node, and is only experimental in the latter case.
96 1 Amber Herold
97
98
Tomography or Exposure/Settings/Image Acquisition>
99
100
101
* Move Type = modeled stage position
102
103
104
* Mover=navigator
105
106
107
* Navigator Target Tolerance = 1 e-7 m (or whatever tolerance you like)
108
109
110
* Navigator Acceptable Tolerance = 1 e-6 m (or whatever tolerance you like)
111
112
113
* Final Image Shift = Yes
114
115
116
117
118
119
120
121
h2. Optimize autofocusing sequence
122
123
124
125 5 Amber Herold
The focus sequence in Focus and Z Focus nodes can and should be optimize for specific cases. The activated sequence in the default setting is good for a flat holey grid and is meant for high resolution imaging at 50k x or higher. If accurate defocus is not important, focusing once per grid square may be sufficient, for which all focus sequence in Focus node can be deactivated and the same autofocus step can be added to Z Focus node instead.
126 1 Amber Herold
127
128 5 Amber Herold
There are two methods to determine the defocus automatically : Stage tilt (Equivalent to stage wobbling) and Beam tilt. There are also two possible ways for correcting the defocus measured: Defocus (Equivalent to turning the focusing knob on the scope and reset the defocus) and Stage Z (Moving the stage to the zero defocus height).
129 1 Amber Herold
130
131 5 Amber Herold
The following observations at NRAMM may help you determine what is the best to use:
132 1 Amber Herold
133
134 5 Amber Herold
*  Both defocus determination method requires calculation correlation between two images. Therefore, the magnification and the location of the autofocusing should include area with contrast.
135 1 Amber Herold
136
137 5 Amber Herold
*  The higher the magnification, the more accurate the defocus determination but the smaller the range of defocus the defocus determination can handle.
138 1 Amber Herold
139
140 5 Amber Herold
*  Beam Tilt is more reproducible and preferred method to determine defocus in HM mode.
141 1 Amber Herold
142
143 5 Amber Herold
*  Stage Tilt (Wobbling) is the preferred method to determine defocus in LM mode.
144 1 Amber Herold
145
146
147 3 Alex Kadokura
148 5 Amber Herold
MSI-Raster Chapter contains an example of an [[Improving Autofocusing|alternative Z Focus node focus sequence]] for grid that show little contrast at medium magnifications.
149 2 Amber Herold
150
151
152
______
153
154
[[Initial MSI application preferences|< Initial MSI application preferences]] | [[MSI Quick-start|MSI Quick-start >]]
155
156
______