Tivips f416 electron diffraction recording » History » Version 4
Anchi Cheng, 12/12/2019 06:48 PM
1 | 1 | Anchi Cheng | h1. TVIPS f416 electron diffraction recording |
---|---|---|---|
2 | |||
3 | TVIPS F416 camera has rolling shutter movie collection capacity, accessible through EMMENU. |
||
4 | |||
5 | h2. Leginon view port configuration setup |
||
6 | |||
7 | This setup will be used by Leginon to control view port 1. The configurations will be changed through EMMNENU scripting |
||
8 | |||
9 | h2. Recorder configuration |
||
10 | |||
11 | h3. Make sure recorder tool (red dot) is shown in toolbar |
||
12 | |||
13 | 3 | Anchi Cheng | Reach toolbar view selection through EMMENU4/View/Toolbar> |
14 | 1 | Anchi Cheng | |
15 | !RecorderToolbar.png! |
||
16 | |||
17 | h3. Set recorder options |
||
18 | |||
19 | Click on the hammer next to the recorder tool to set these. |
||
20 | |||
21 | Basic settings |
||
22 | !RecorderBasicSettings.png! |
||
23 | |||
24 | 2 | Anchi Cheng | Advanced settings: save in tvips image set format |
25 | 1 | Anchi Cheng | !RecorderAdvancedSettings.png! |
26 | 4 | Anchi Cheng | |
27 | h2. Leginon instruments.cfg |
||
28 | |||
29 | Replace |
||
30 | <pre> |
||
31 | class:tietz.TietzF416 |
||
32 | </pre> |
||
33 | with |
||
34 | <pre> |
||
35 | clas:tietz2.EmMenuF416 |
||
36 | </pre> |
||
37 | |||
38 | h2. Set up tvips.cfg |
||
39 | |||
40 | # Copy pyscope/tvips.cfg.template to the same directory where instruments.cfg is read on TVIPS PC and rename it as *tvips.cfg* |
||
41 | # Match the recorder directory and default filename with what is in EMMENU4 recorder settings. |