h1. TVIPS f416 electron diffraction recording
TVIPS F416 camera has rolling shutter movie collection capacity, accessible through EMMENU.
h2. Leginon view port configuration setup
This setup will be used by Leginon to control view port 1. The configurations will be changed through EMMNENU scripting
h2. Recorder configuration
h3. Make sure recorder tool (red dot) is shown in toolbar
Reach toolbar view selection through EMMENU4/View/Toolbar>
!RecorderToolbar.png!
h3. Set recorder options
Click on the hammer next to the recorder tool to set these.
If you have installed an SSD, set it to a directory on that drive which can speed up the saving.
Basic settings
!RecorderBasicSettings.png!
Advanced settings: save in tvips image set format
!RecorderAdvancedSettings.png!
h2. Leginon instruments.cfg
Replace
<pre>
class:tietz.TietzF416
</pre>
with
<pre>
clas:tietz2.EmMenuF416
</pre>
h2. Set up tvips.cfg
# Copy pyscope/tvips.cfg.template to the same directory where instruments.cfg is read on TVIPS PC and rename it as *tvips.cfg*
# Match the recorder directory and default filename with what is in EMMENU4 recorder settings.