Tivips f416 electron diffraction recording » History » Version 6
Anchi Cheng, 07/29/2021 12:46 PM
1 | 1 | Anchi Cheng | h1. TVIPS f416 electron diffraction recording |
---|---|---|---|
2 | |||
3 | TVIPS F416 camera has rolling shutter movie collection capacity, accessible through EMMENU. |
||
4 | |||
5 | h2. Leginon view port configuration setup |
||
6 | |||
7 | 6 | Anchi Cheng | [[Create Leginon view port]] |
8 | |||
9 | 1 | Anchi Cheng | This setup will be used by Leginon to control view port 1. The configurations will be changed through EMMNENU scripting |
10 | |||
11 | h2. Recorder configuration |
||
12 | |||
13 | h3. Make sure recorder tool (red dot) is shown in toolbar |
||
14 | |||
15 | 3 | Anchi Cheng | Reach toolbar view selection through EMMENU4/View/Toolbar> |
16 | 1 | Anchi Cheng | |
17 | !RecorderToolbar.png! |
||
18 | |||
19 | h3. Set recorder options |
||
20 | |||
21 | Click on the hammer next to the recorder tool to set these. |
||
22 | 5 | Anchi Cheng | If you have installed an SSD, set it to a directory on that drive which can speed up the saving. |
23 | 1 | Anchi Cheng | |
24 | Basic settings |
||
25 | !RecorderBasicSettings.png! |
||
26 | |||
27 | 2 | Anchi Cheng | Advanced settings: save in tvips image set format |
28 | 1 | Anchi Cheng | !RecorderAdvancedSettings.png! |
29 | 4 | Anchi Cheng | |
30 | h2. Leginon instruments.cfg |
||
31 | |||
32 | Replace |
||
33 | <pre> |
||
34 | class:tietz.TietzF416 |
||
35 | </pre> |
||
36 | with |
||
37 | <pre> |
||
38 | clas:tietz2.EmMenuF416 |
||
39 | </pre> |
||
40 | |||
41 | h2. Set up tvips.cfg |
||
42 | |||
43 | # Copy pyscope/tvips.cfg.template to the same directory where instruments.cfg is read on TVIPS PC and rename it as *tvips.cfg* |
||
44 | # Match the recorder directory and default filename with what is in EMMENU4 recorder settings. |