1 | 4 | Anchi Cheng | h1. Using Gatan K2 Summit in Leginon |
---|---|---|---|
2 | 1 | Anchi Cheng | |
3 | 4 | Anchi Cheng | The main use of K2 Summit is in Counted or Super-resolution mode in combination with dose fractionation (frame saving). |
4 | |||
5 | 1 | Anchi Cheng | If you want to use DM's dose fractionation, activate the "save frames" check box in the particular preset camera configuration. You can also use DM's frame alignment algorithm by activating "align frames" check box, but the latter is not recommended since the alignment in DM will delay the return of the image to Leginon and therefore affects throughput greatly. |
6 | |||
7 | 27 | Anchi Cheng | * [[K2 Summit introduction]] |
8 | * [[K2 acquire image|Acquiring summed frame image thru Leginon]] |
||
9 | * [[K2 gain correction|Gain and Dark Correction]] *This involves both DM and Leginon* |
||
10 | * [[K2 calibration|Pixel and Matrix Calibrations]] |
||
11 | * [[K2 raw frame saving|Trigger raw frame saving]] |
||
12 | * [[DDD raw frame file transfer|Transferring raw frame images]] off the hard drive of K2 Summit during data collection to a network drive. |
||
13 | * [[Make DDD movie stack|Compiling corrected movie stack]] with Appion using bright and dark images acquired in Leginon |
||
14 | 2 | Anchi Cheng | |
15 | 9 | Anchi Cheng | h2. Warning |
16 | |||
17 | 19 | Anchi Cheng | * Because of a delay in communication with DM when Leginon is started, camera configuration shown in Correction Node settings may not corresponds to what is used last time when you quit Leginon. Make sure you check this. |
18 | 6 | Anchi Cheng | |
19 | h2. Leginon functions that do not work with K2 camera |
||
20 | |||
21 | h3. General functions |
||
22 | # The dose matching tool in Preset Manager. K2's exposure time is not fine grained enough to make small adjustment according to the small dose deviation. You will have to use fixed exposure time and adjust beam intensity by hand to achieve the right dose. We find out the equivalent screen reading from the microscope for the acceptable 8-10 e/physical pixel/s dose rate for the detector and make adjustment there. |
||
23 | 20 | Anchi Cheng | # Tomography correlation algorithm only takes square images. Until that is changed, Frame saving is not possible since only a full size unbinned image can be paired with frame saving. |
24 | 6 | Anchi Cheng | |
25 | h3. Frame saving camera functions |
||
26 | # The image returned to Leginon display is always integrated over all the frames with K2 summit camera. Selection of frames to use as in DE series does not work here. |
||
27 | # Readout delay is not adjustable on K2. Leave it at 0. |
||
28 | 2 | Anchi Cheng | |
29 | h2. Clean up |
||
30 | |||
31 | Frame saving generates a lot of files during acquisition and during processing. |
||
32 | |||