Bug #1202
closed
- Category set to myamiweb
- Status changed from Assigned to In Code Review
- Assignee changed from Anchi Cheng to Amber Herold
- Target version set to Appion/Leginon 2.2.0
r15417 fix this. Data array is in meters for consistency but show drift rate in graph in Angstrom for convention.
Test:
1. Use a session you are currently running. Finde drift rate measured by Drift Monitor, note the number.
2. Go to myamiweb summary of the session (click on the top left part of an image viewer where it says summary).
3. Click on the graph for drift and see if the last point of the data corresponds to the measuement logged in Leginon.
Any Leginon MSI user can test.
- Assignee changed from Amber Herold to Anchi Cheng
looks good. I would suggest comments in the code for the magic numbers 1.63e-10 * 4 and 1e-10 to explain what they are and why they are being used.
- Assignee changed from Anchi Cheng to Amber Herold
on second thought, may as well make it to display in pixel if the query for pixelsize fail, save explanation of what that value from tecnai2 is. r15473
- Status changed from In Code Review to In Test
- Assignee changed from Amber Herold to Dmitry Lyumkis
Looks good, ready for test.
- Status changed from In Test to Closed
Also available in: Atom
PDF