Actions
Bug #2659
closedBeam stays tilted when low level error occurs in BeamTiltCalibrationClient
Status:
Closed
Priority:
High
Assignee:
Category:
-
Target version:
Start date:
02/15/2014
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 2.1.0
Show in known bugs:
No
Workaround:
Description
Even though the beam should be reset back to rotation center, some error causes an early return of the function measureDefocusStig.
An example is when tilt correction put the image outside the image and resulted in a flat image.
This is found in an RCT session where "RCT Focus" node move type was set to image shift and is a serious bug since it is hard for user to realize what the problem is.