Bug #4794
closed
  
    
    
  
Need more than 3 values to unpack error during data collection
 
        
        Added by Ed Eng over 8 years ago.
        Updated almost 8 years ago.
        
Affected Version:
Appion/Leginon 2.2.0
 
 
  
  
  
  Description
  
  2017/02/28
Collecting at -30 degree tilt
Session: 17feb28b
Priyamvada encountered a few errors:
	
	- preset request to "hln" failed, waiting 60 seconds to try again
 
		- Move to target failed: unable to set instrument
 
		- need more than 3 values to unpack
 
	
	No errors within Leginon clients. Sequenced paused within K1n_Z_Focus sequence at hln step.
	Restarted Leginon and continuing data collection.
   
 
  
  Files
  
 
  
  
    
    
    
    
       - Status changed from New to Assigned
 
    
    lost communication to scope.  may need to add a retry.
 
     
   
  
  
    
    
    
    unpack error may be in pyscope/tecnai.py _setStagePosition function unpacking exc.  Need to check on scope
 
     
   
  
  
    
    
    
    
       - Project changed from 138 to Leginon
 
       - Subject changed from Leginon error during Krios1 tilt data collection to Need more than 3 values to unpack error during data collection
 
       - Affected Version set to Appion/Leginon 2.2.0
 
    
    The root problem is lost of communication to COM, resulting a COMError.  The exception raised do not seem to have a fixed format or has been changed since the pyscope code handling it was written.
 
     
   
  
  
    
    
    
    I can confirm that for the past 3-4 months this error has been showing up during tilt-series collection on Krios1 and has increased with time. I don't think I've seen it on Krios2, and I'm pretty sure I haven't seen it on non-Krios scopes. When Krios1 gets a PM this should be brought up.
 
     
   
  
  
    
    
    
    
    2017/03/08
Occurred multiple times during session for rhite during alpha tilt.
 
     
   
  
  
    
    
    
    I will make more permanent log on position where this happens on the scope computer.
 
     
   
  
  
    
    
    
    
       - Related to Bug #4951: Access Denied ComError in stage movement on Krios added
 
    
    
     
   
  
  
    
    
    
    
       - Status changed from Assigned to Closed
 
    
    
     
   
  
 
  
  
 
Also available in:  Atom