Actions
Feature #7191
openalign_ZLP node to send warning to slack when slit adjustment >2 eV
Start date:
04/10/2019
Due date:
% Done:
0%
Estimated time:
Deliverable:
Description
Can we get a feature in leginon for krios2 and krios3 where the align_zlp node will send a slack warning when the slit adjustment is greater than some value, say > 2 eV? Something to warn us that there is potentially some instability in the system (i.e. temperature changes in the room). This would be helpful since recently on 4/9/19 there was a change in our HVAC system that caused a warm up in the control room from 75.5 F to 78.5 F from 3AM to 6AM. This drastic change in temp caused the slit to become unstable, and slipping into many of the images on krios2 and krios3.
Updated by Anchi Cheng about 4 years ago
- Target version set to Appion/Leginon Future Version