Project

General

Profile

Actions

Feature #7191

open

align_ZLP node to send warning to slack when slit adjustment >2 eV

Added by Laura Yen over 5 years ago. Updated about 4 years ago.

Status:
New
Priority:
Normal
Assignee:
-
Category:
-
Start date:
04/10/2019
Due date:
% Done:

0%

Estimated time:
Deliverable:

Description

Can we get a feature in leginon for krios2 and krios3 where the align_zlp node will send a slack warning when the slit adjustment is greater than some value, say > 2 eV? Something to warn us that there is potentially some instability in the system (i.e. temperature changes in the room). This would be helpful since recently on 4/9/19 there was a change in our HVAC system that caused a warm up in the control room from 75.5 F to 78.5 F from 3AM to 6AM. This drastic change in temp caused the slit to become unstable, and slipping into many of the images on krios2 and krios3.

Actions

Also available in: Atom PDF