Bug #739
closed
When RCT will collects a zero degree image it sets targets to complete and does not collect other tilts
Added by Neil Voss over 14 years ago.
Updated about 14 years ago.
Category:
Leginon subsystem
Description
Doing RCT with beta leginon, I select 3 targets and 1 focus spot. Leginon tilts, focuses and collects the 3 images. Then it tilts to second tilt angle, complains that there is no focus spot and that all images have been collected.
Files
- Assignee deleted (
Jim Pulokas)
This might be a the bug (that I thought was fixed) where your first tilt is zero it automatically fails all subsequent tilts.
- Subject changed from RCT does focus, collect images for first tilt then on second tilt complains there is no focus target to When RCT will collects a zero degree image it sets targets to complete and does not collect other tilts
- Status changed from New to Assigned
- Assignee set to Neil Voss
- Priority changed from Urgent to Normal
- Target version changed from Appion/Leginon 2.0.2 to Appion/Leginon 2.1.0
- Affected Version changed from Appion/Leginon 2.0.1 to Pre-2.0
- Status changed from Assigned to In Code Review
- Assignee changed from Neil Voss to Anchi Cheng
- % Done changed from 0 to 100
- Status changed from In Code Review to In Test
- Assignee changed from Anchi Cheng to Arne Moeller
Has anyone looked into this bug? Lots of people here are complaining that they can no longer collect tilted data.
I ran RCT last Friday and it seemed fine to me.
- Status changed from In Test to Closed
- Assignee deleted (
Arne Moeller)
Yeah, this is fixed in 2.0, the bug is marked as 'in test'. This bug applied to only small angle tilts, because I had radians instead of degrees and it screwed things up with the small angles, so you should not have this bug Gabe.
I am closing this bug cause Arne says it works.
Gabe,
Since I merged the fix to 2.0 branch 17 days ago, you might want to do an svn update. Don't know if Amber rolled out another revision after my merge.
Anchi
Also available in: Atom
PDF