Project

General

Profile

Actions

Bug #7834

closed

Shutter control during acquisition does not work if pre-specimen shutter is set to Shutter number 2 in GMS

Added by Anchi Cheng over 5 years ago. Updated almost 5 years ago.

Status:
Closed
Priority:
High
Assignee:
Category:
-
Target version:
Start date:
07/22/2019
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 3.4
Show in known bugs:
No
Workaround:

Description

Found this problem with GMS 3.22.1460

Actions #1

Updated by Anchi Cheng over 5 years ago

This bug may not have been seen before because of a bug in the earlier GMS. GMS shutter configuration varies greatly.

The working configuration on TFS scope is this:

Acquisition shutter (Pre-specimen) should be set to

Configuration=Open/5V(inverted)
Idle State=Closed

Post-speciment should be set to

Configuration=Open/5V(inverted)
Idle State=Open
Actions #2

Updated by Anchi Cheng almost 5 years ago

  • Status changed from Assigned to Closed
  • Target version set to Appion/Leginon 3.5