Falcon acquire image » History » Version 5

Anchi Cheng, 08/08/2014 11:43 AM

1 1 Anchi Cheng
h1. Acquire images on Falcon II
2
3
h2. Preparation in TUI/TIA
4
5
h3. Acquire Gain/Dark References
6
7 5 Anchi Cheng
Follow [[Falcon gain correction]] section
8 3 Anchi Cheng
9 1 Anchi Cheng
h2. Recommended dose rate
10
11
The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera.
12 3 Anchi Cheng
If you have to use lower dose rate, check the validity of the gain reference.
13 1 Anchi Cheng
14
h2. Preset Recommendation
15
16
An example can be found in [[Pre-MSI_Set-up#Preset-Design-Example-for-Falcon-II-on-Krios]]