Falcon acquire image » History » Version 5
Anchi Cheng, 08/08/2014 11:43 AM
| 1 | 1 | Anchi Cheng | h1. Acquire images on Falcon II |
|---|---|---|---|
| 2 | |||
| 3 | h2. Preparation in TUI/TIA |
||
| 4 | |||
| 5 | h3. Acquire Gain/Dark References |
||
| 6 | |||
| 7 | 5 | Anchi Cheng | Follow [[Falcon gain correction]] section |
| 8 | 3 | Anchi Cheng | |
| 9 | 1 | Anchi Cheng | h2. Recommended dose rate |
| 10 | |||
| 11 | The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera. |
||
| 12 | 3 | Anchi Cheng | If you have to use lower dose rate, check the validity of the gain reference. |
| 13 | 1 | Anchi Cheng | |
| 14 | h2. Preset Recommendation |
||