Faraday Cup Measurment Notes¶
Picoammeter setup:- plug in
- wait 1 hour
- activate ZCHK (maybe already when turned on) (should say ZC on screen)
- plug into faraday holder
- set to lowest range (2nA)
- activate ZCOR to set zero (should say ZZ on screen)
- set auto range if desired
- deactivate ZCHK to begin measuring
- 1.00 nA
Exposure Time to Screen Current measurment¶
- refer to Tecnai Help for "Exposure Time, Plate Camera CP"
- formula: beam current (nA) = 2.15 * emulsion / measured_exposure_time
- exp time is between 3.1 and 3.5 depending on size of beam
- emulsion = 1
- beam current is between 0.61 and 0.69 nA