 Picoammeter setup: 
 * plug in 
 * wait 1 hour 
 * activate ZCHK (maybe already when turned on) (should say ZC on screen) 
 * plug into faraday holder 
 * set to lowest range (2nA) 
 * activate ZCOR to set zero    (should say ZZ on screen) 
 * set auto range if desired 
 * deactivate ZCHK to begin measuring 
 ** 1.00 nA 


 h1. Exposure Time to Screen Current measurment 

 * refer to Tecnai Help for "Exposure Time, Plate Camera CP" 
 * formula:    beam current (nA) = 2.15 * emulsion / measured_exposure_time 
 ** exp time is between 3.1 and 3.5 depending on size of beam 
 ** emulsion = 1 
 ** beam current is between 0.61 and 0.69 nA