Project

General

Profile

Faraday Cup Measurment Notes » History » Version 2

Jim Pulokas, 10/04/2010 03:30 PM

1 1 Jim Pulokas
h1. Faraday Cup Measurment Notes
2
3
Picoammeter setup:
4
* plug in
5
* wait 1 hour
6
* activate ZCHK (maybe already when turned on) (should say ZC on screen)
7
* plug into faraday holder
8
* set to lowest range (2nA)
9
* activate ZCOR to set zero  (should say ZZ on screen)
10
* set auto range if desired
11
* deactivate ZCHK to begin measuring
12 2 Jim Pulokas
** 1.00 nA
13
14
15
h1. Exposure Time to Screen Current measurment
16
17
* refer to Tecnai Help for "Exposure Time, Plate Camera CP"
18
* formula:  beam current (nA) = 2.15 * emulsion / measured_exposure_time
19
** exp time is between 3.1 and 3.5 depending on size of beam
20
** emulsion = 1
21
** beam current is between 0.61 and 0.69 nA