1 | 1 | Jim Pulokas | h1. Faraday Cup Measurment Notes |
---|---|---|---|
2 | |||
3 | Picoammeter setup: |
||
4 | * plug in |
||
5 | * wait 1 hour |
||
6 | * activate ZCHK (maybe already when turned on) (should say ZC on screen) |
||
7 | * plug into faraday holder |
||
8 | * set to lowest range (2nA) |
||
9 | * activate ZCOR to set zero (should say ZZ on screen) |
||
10 | * set auto range if desired |
||
11 | * deactivate ZCHK to begin measuring |
||
12 | 2 | Jim Pulokas | ** 1.00 nA |
13 | |||
14 | |||
15 | h1. Exposure Time to Screen Current measurment |
||
16 | |||
17 | * refer to Tecnai Help for "Exposure Time, Plate Camera CP" |
||
18 | * formula: beam current (nA) = 2.15 * emulsion / measured_exposure_time |
||
19 | ** exp time is between 3.1 and 3.5 depending on size of beam |
||
20 | ** emulsion = 1 |
||
21 | ** beam current is between 0.61 and 0.69 nA |