1 | 1 | Jim Pulokas | h1. Faraday Cup Measurment Notes |
---|---|---|---|
2 | |||
3 | Picoammeter setup: |
||
4 | * plug in |
||
5 | * wait 1 hour |
||
6 | * activate ZCHK (maybe already when turned on) (should say ZC on screen) |
||
7 | * plug into faraday holder |
||
8 | * set to lowest range (2nA) |
||
9 | * activate ZCOR to set zero (should say ZZ on screen) |
||
10 | * set auto range if desired |
||
11 | * deactivate ZCHK to begin measuring |
||
12 | 3 | Jim Pulokas | * focus beam to a spot |
13 | * pull out holder until spot hits the faraday cup (max current) |
||
14 | ** 200keV |
||
15 | *** 1.01 nA |
||
16 | ** 120 keV |
||
17 | *** 2.53 nA |
||
18 | 2 | Jim Pulokas | |
19 | |||
20 | h1. Exposure Time to Screen Current measurment |
||
21 | |||
22 | * refer to Tecnai Help for "Exposure Time, Plate Camera CP" |
||
23 | * formula: beam current (nA) = 2.15 * emulsion / measured_exposure_time |
||
24 | ** exp time is between 3.1 and 3.5 depending on size of beam |
||
25 | ** emulsion = 1 |
||
26 | 1 | Jim Pulokas | ** beam current is between 0.61 and 0.69 nA |
27 | 3 | Jim Pulokas | |
28 | |||
29 | h1. Leginon screen scale factor measurement |
||
30 | |||
31 | * adjust scale factor until beam current is same as faraday cup |
||
32 | ** 200 keV |
||
33 | *** 0.725 |
||
34 | ** 120 keV |
||
35 | |||
36 | h1. Camera sensitivity |
||
37 | |||
38 | ** 200 keV |
||
39 | *** 45-50 count/e |
||