h1. Faraday Cup Measurment Notes
2
3
Picoammeter setup:
4
* plug in
5
* wait 1 hour
6
* activate ZCHK (maybe already when turned on) (should say ZC on screen)
7
* plug into faraday holder
8
* set to lowest range (2nA)
9
* activate ZCOR to set zero  (should say ZZ on screen)
10
* set auto range if desired
11
* deactivate ZCHK to begin measuring
12 3 Jim Pulokas
* focus beam to a spot
13
* pull out holder until spot hits the faraday cup (max current)
14
** 200keV
15
*** 1.01 nA
16
** 120 keV
17
*** 2.53 nA
18 2 Jim Pulokas
19
20
h1. Exposure Time to Screen Current measurment
21
22
* refer to Tecnai Help for "Exposure Time, Plate Camera CP"
23
* formula:  beam current (nA) = 2.15 * emulsion / measured_exposure_time
24
** exp time is between 3.1 and 3.5 depending on size of beam
25
** emulsion = 1
26 1 Jim Pulokas
** beam current is between 0.61 and 0.69 nA
27 3 Jim Pulokas
28
29
h1. Leginon screen scale factor measurement
30
31
* adjust scale factor until beam current is same as faraday cup
32
** 200 keV
33
*** 0.725
34
** 120 keV
35
36
h1. Camera sensitivity
37
38
** 200 keV
39
*** 45-50 count/e
40
** 120 keV
41
*** 107 count/e