Project

General

Profile

Gatan K2 Instruments cfg » History » Version 5

Anchi Cheng, 01/07/2014 10:08 PM

1 4 Anchi Cheng
The camera modes are configured as separate cameras as follows.    
2 1 Anchi Cheng
3 4 Anchi Cheng
* Super Resolution should be double the size of Linear and Counting.  
4
* zplane will be relative to any other cameras in your system.  Anything below K2 should have a lower zplane.
5 5 Anchi Cheng
* to make the images acquired from the camera to have the [[Leginon Image Orientation|standard Leginon orientation]], the camera configuration in Digital Micrograph will need a rotation and flip (270 degree rotation and a flip around the vertical axis on our camera).  Therefore, *the height is longer than width* in the configuration below
6 2 Anchi Cheng
7 1 Anchi Cheng
<pre>[Gatan K2 Linear]
8
class: dmsem.GatanK2Linear
9
zplane: 50
10
width: 3710
11
height: 3838
12
13
[Gatan K2 Counting]
14
class: dmsem.GatanK2Counting
15
zplane: 50
16
width: 3710
17
height: 3838
18
19
[Gatan K2 Super]
20
class: dmsem.GatanK2Super
21
zplane: 50
22
width: 7420
23
height: 7676</pre>
24 4 Anchi Cheng
25
If you have an Orius camera with K2, set its zplane below K2 like this:
26
<pre>
27
[camera2]
28
class: dmsem.GatanOrius
29
zplane: 49
30
</pre>