Because MSI-raster is normally used for acquiring images on negatively stained particles on continuous carbon, the intermediate mag preset (hl) often lacks contrast. To reduce focusing failure, we recommend a different focusing sequence for the Z focus node.
5
6
7
8
* Use Stage Alpha Tilt to determine defocus at the low magnification sq preset to do a rough stage-z adjustment to U-center
9
10
11
* Refine stage-z position at the high magnification fa preset and Use beam-tilt to determine the required correction.
12
13
14
15
16
17
For "Stage_Wobble" Step:
18
*  Enable = Yes
19
20
21
*  Image registration= phase correlation
22
23
24
*  Fit limit = 1000 (not used)
25
26
27
*  Correct for delta Defocus/Z between 0 and 2e-4 meters
28
29
30
*  Correction type = Stage Z
31
32
33
*  Check drift greater than 3e-10 meters/s = No
34
35
36
*  Stigmator correction = No
37
38
39
*  Stigmator Defocus Max/Min = N/A
40
41
42
43
44
45
For "Z_to_Eucentric" Step, change these:
46
*  preset = fa
47
48
49
*  Correct for delta Defocus/Z between 0 and 2e-5 meters
50 2 Amber Herold