Project

General

Profile

Krios1-K2-NYSBC-large shift » History » Version 3

Anchi Cheng, 03/21/2019 06:43 PM

1 1 Anchi Cheng
h1. Krios1-K2-NYSBC-large shift
2
3
h1. Krios-K2 Presets used for 1.9 A mouse recombinant apoferritin reconstruction data collection at SEMC @ NYSBC
4
5
The same as [[Krios1-K2-NYSBC]] except
6
7
|Probe|_.Lens Series|Magnification:|_.Preset name:|_.Image Shift (x,y):|_.Dimension:|_.Binning:|_.Camera Dose (e/camera pixel/s):|_.Exposure Time (ms):|_.Specimen Dose Rate (e/A^2):|_.Spot Size:|_.Defocus (m):|_.C2 (um):|frame saving:|Camera|
8
|nano|SA|1300**|hl|Aligned|927x959|4|8|200|0.004|11|-1.5e-4|70|no|K2|
9 2 Anchi Cheng
|nano|SA|37000|en|0,0|3710x3838|1|4.2 e/pixel/s and ~1.12 um on the specimen|6000|58|6|(-1e-6 to -2e-6)|70|200 ms/frame|K2|
10 1 Anchi Cheng
11 3 Anchi Cheng
**This magnification is an add-on to standard SA lens series made for NYSBC Krios1. May not be available at other scopes.
12 1 Anchi Cheng
13
* Note: no defocal pair was acquired in this experiment.  Therefore, no ef preset.