Wim Hagen
- Login: wimhagen
- Email: hagen@embl.de
- Institution: EMBL Heidelberg
- Registered on: 05/16/2016
- Last connection: 07/07/2016
Issues
open | closed | Total | |
---|---|---|---|
Assigned issues | 0 | 0 | 0 |
Reported issues | 0 | 0 | 0 |
Activity
03/27/2022
- 03:28 PM Leginon Leginon: Error using EF service camera
- Hi Trishant,
To my knowledge, this camera is only available for customer service purposes, and only on the earlier S...
03/07/2022
- 11:02 AM Leginon Leginon: TEM Controller options
- Apertures are only accessible with some special scripting option that nobody has ever been able to obtain. Could be t...
11/17/2021
- 12:32 PM Leginon Leginon: RE: Progress on Selectris support?
- Keep in mind the EFTEM lens series for Selectris systems is software version dependent, maybe that causes trouble?
Be...
10/11/2021
- 07:43 AM Leginon Leginon: Falcon 4 error with getImage()
- Hi Morgan,
My attachment didn’t get through:
https://oc.embl.de/index.php/s/KhsUwzFwu39PiX8 <https://oc.embl.de/ind... - 07:39 AM Leginon Leginon: Falcon 4 error with getImage()
- Hi mriagn,
Run c:/tecnai/exe/service/LicenseReader.exe, to check if TFS have actually updated your microscope’s cust...
09/13/2021
- 06:23 AM Leginon Leginon: Progress on Selectris support?
- Something to be aware of:
As of TFS temserver SW version 7.9.0 (so 3.9.0 for Krios and 2.9.0 for Talos/Glacios), the ...
07/07/2016
- 12:49 PM Leginon Leginon: RE: Stage seems not to always return to zero degrees upon completion of a tilt series
- Keep in mind that there is no feedback loop for stage tilt after a tilt change, FEI software does not act on e.g. tra...
06/22/2016
- 12:22 PM Leginon Leginon: RE: Acquire seamless images for quantitative TEM
- Hi Claudio,
On an FEI TEM, at the magnification you need for a 7 micron tile size, you will only have about 25 micro...
05/16/2016
- Hi all,
FEI's direct alignment does not allow coma-free direct alignment with a beam-image shift applied, it will ...