- 08:27 PM Leginon Leginon: RE: Error acquiring image: Error sending request: Request
- Well, now there are errors at different places.
Here is a screenshot of the error log and the client log after runn... - 08:05 PM Leginon Leginon: RE: Error acquiring image: Error sending request: Request
- with save frames
- 08:01 PM Leginon Leginon: RE: Error acquiring image: Error sending request: Request
- Oh yes, specifically, it will take
Raw, Channel 0, Acquire = no error
Dark, Both Channels, Acquire = error
then ... - 07:37 PM Leginon Leginon: RE: Error acquiring image: Error sending request: Request
- Hi Anchi,
Thank you so much. The pyscope test is attached below. Let me know if you need anything else to help di... - Hi,
We have been using leginon 3.5 for the last 6 months and things were working until I updated SerialEM on our ...
05/10/2021
- 01:04 PM Leginon Leginon: RE: K3 sensitivity value very low
- Yes great! Thank you so much, Anchi. The image looked very dark before. Now the sensitivity is 0.969 just like it sho...
05/07/2021
- 11:10 PM Leginon Leginon: RE: K3 sensitivity value very low
- Sorry, yes I mean the ratio between leginon counts/GMS electrons. It's very low when I use either CDS or when I use n...
- 08:40 PM Leginon Leginon: RE: K3 sensitivity value very low
- OK thank you! If it helps, I am having the same sensitivity/ratio issue with CDS and without.
- 01:08 PM Leginon Leginon: RE: K3 sensitivity value very low
- I am in the Navigation node and I sent my en preset to do the reading (it is 8eps in GMS at GMS bin1). The mean count...
05/06/2021
- While doing dose calibration (camera sensitivity) for our Glacios/K3 I am getting an extremely low number that I don'...