closedeucentric point determination by stage tilt does not work well when there is big jump in tilt direction change
Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
04/10/2014
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 2.1.0
Show in known bugs:
No
Workaround:
Description
Only the end points of the tilt are used for this measurement. Therefore, the measurement is over estimated.
This was observed on Krios.
Updated by Anchi Cheng over 10 years ago
r18167 averages the two results from the two tilt directions.
Updated by Anchi Cheng over 10 years ago
- Related to Bug #2683: measurement of stage z from Eucentric height is wrong when tilt backlash is significant added
Updated by Anchi Cheng over 8 years ago
- Status changed from Assigned to Closed
- Target version set to Appion/Leginon 3.1.0
This was fixed a while ago.