eucentric point determination by stage tilt does not work well when there is big jump in tilt direction change

Added by Anchi Cheng over 10 years ago. Updated over 8 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
04/10/2014
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.1.0
Show in known bugs:
No
Workaround:

Description

Only the end points of the tilt are used for this measurement. Therefore, the measurement is over estimated.

This was observed on Krios.


Related issues 1 (0 open1 closed)

Related to Leginon - Bug #2683: measurement of stage z from Eucentric height is wrong when tilt backlash is significantClosedAnchi Cheng03/07/2014