Actions
Bug #562
closedFocus sequence problem
Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
05/25/2010
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 2.0.1
Show in known bugs:
Yes
Workaround:
Give 2 different names for different sequences.
Description
If you have 2 steps in focus sequence at the same preset (i.e. repeat for better convergence), defocus measurement fails because defocus is not applied for the second stage.
Preset does not change for the second step.
Fix: Force always a preset change in focus sequence.
Updated by Neil Voss over 14 years ago
- Target version set to Appion/Leginon 2.1.0
- Affected Version set to Appion/Leginon 2.0.1
Updated by Neil Voss over 14 years ago
- Status changed from New to Assigned
- Assignee set to Anchi Cheng
Updated by Anchi Cheng over 14 years ago
- Assignee changed from Anchi Cheng to Jim Pulokas
- Show in known bugs set to No
Jim,
You'd better look at this. I thought of just set instrument defocus but not sure if there are other consequences such as if it is tilted and image shifted.
I will test.
Updated by Amber Herold about 14 years ago
- Target version changed from Appion/Leginon 2.1.0 to Appion/Leginon 2.1.1
- Show in known bugs changed from No to Yes
- Workaround updated (diff)
This fix will involve much code making it a risky change this close to release. It will be fixed for the next release.
Updated by Jim Pulokas about 11 years ago
- Status changed from Assigned to Closed
fixed in related issue
Actions