1 | 6 | Anchi Cheng | h1. Continuous tilt electron diffraction recording use Ceta or Ceta-D camera and TFS scopes |
---|---|---|---|
2 | 1 | Anchi Cheng | |
3 | Micro-ED involves recording 3D crystal electron diffraction pattern during a continuous stage tilt. Leginon implementation of this application includes the listed components below and is available by using myami-ed branch until it is officially added to future release. As is, it is only fully implemented for TFS Ceta or Ceta-D camera. |
||
4 | |||
5 | Since many of the feature required here are not available through Standard or Advanced TEM Scripting from TFS, AutoIt scripts are used in several places. |
||
6 | |||
7 | h2. Components of the feature |
||
8 | |||
9 | 5 | Anchi Cheng | # Diffraction mode TEM instrument |
10 | 8 | Anchi Cheng | # Beamstop control: achieved by AutoIt Scripts BeamstopIn and BeamstopOut |
11 | 9 | Anchi Cheng | # Rolling-shutter movie acquisition using TIA interface |
12 | # Data conversion and upload of the movies into Leginon database and crystallography format |
||
13 | 4 | Anchi Cheng | # [[ MSI-Diffr application ]] |
14 | 11 | Anchi Cheng | # [[Camera length calibration]] |
15 | 1 | Anchi Cheng | |
16 | h2. Installation (assuming that you've already have Leginon installation). |
||
17 | |||
18 | # Check that you have access to Advanced TEM Scripting (frame-saving upgrade not needed). |
||
19 | 12 | Anchi Cheng | # Install or set environment to use git branch myami-ed or current myami-beta on TFS microscope with Ceta camera as well as your leginon linux box. |
20 | 10 | Anchi Cheng | # import updated Calibrations application (see [[Steps_involved_in_the_installation]]) |
21 | # import MSI-diffraction and settings from your_myami/leginon/applications (see general description of [[Steps_involved_in_the_installation]] regarding importing application and additional settings for the application) |
||
22 | 5 | Anchi Cheng | # [[Setup Diffraction mode TEM instrument]] |
23 | 7 | Anchi Cheng | # [[Setup Ceta to use Advanced TEM Scripting]] |
24 | 3 | Anchi Cheng | # [[AutoIt program and script compilation]] |
25 | 2 | Anchi Cheng | |
26 | h2. Usage |
||
27 | |||
28 | 15 | Anchi Cheng | h3. Check these before you start: |
29 | 1 | Anchi Cheng | |
30 | 15 | Anchi Cheng | 1. Confirm that instruments.cfg is properly set up to use feicam for Ceta camera. Especially if you use Falcon camera on the same instrument for imaging. |
31 | 2. Open TUI. Select and insert BM-Ceta in the camera tab |
||
32 | 3. Make sure TIA window is available and the Export Series shortcut is shown. |
||
33 | |||
34 | 13 | Anchi Cheng | h3. Start MSI-DIffr application and assign client as required. |
35 | |||
36 | 1 | Anchi Cheng | h3. [[Setup df preset]] |
37 | 15 | Anchi Cheng | |
38 | h3. Grid atlas collection |
||
39 | |||
40 | 1. Select gr preset to scope, with flucam showing the grid and aperture, draw on the computer screen the location of the centerred small C2 aperture. |
||
41 | 2. Select a larger C2 aperture for grid atlas collection. |
||
42 | 3. collect grid atlas. |
||
43 | 4. Reselect the small C2 aperture. Make sure it is still centered. |
||
44 | |||
45 | h3. Queue up potential crystals |
||
46 | |||
47 | 1. Pick square and then pick "hole" for intermediate mag image of potential crystals. (This is no different from any other MSI application). |
||
48 | 2. Submit preview targets in "DExposure Targeting" node for a small rotation image and confirm whether it diffracts or not. |
||
49 | 3. Submit the good crystal positions as acquisition targets into the queue in "DExposure Targeting" node. No focus target is needed. |
||
50 | 4. If needed, modify "DExposure" settings for start, range, and speed of the tilt. |
||
51 | 5. Click "process queue" tool to start collection |