Faraday Cup Measurment Notes » History » Version 11
Anchi Cheng, 04/18/2013 07:12 PM
1 | 1 | Jim Pulokas | h1. Faraday Cup Measurment Notes |
---|---|---|---|
2 | |||
3 | Picoammeter setup: |
||
4 | * plug in |
||
5 | * wait 1 hour |
||
6 | * activate ZCHK (maybe already when turned on) (should say ZC on screen) |
||
7 | * plug into faraday holder |
||
8 | * set to lowest range (2nA) |
||
9 | * activate ZCOR to set zero (should say ZZ on screen) |
||
10 | * set auto range if desired |
||
11 | * deactivate ZCHK to begin measuring |
||
12 | 3 | Jim Pulokas | * focus beam to a spot |
13 | 11 | Anchi Cheng | * pull out holder until spot hits the faraday cup (max current) (This is a negative value) |
14 | Examples |
||
15 | 3 | Jim Pulokas | ** 200keV |
16 | 11 | Anchi Cheng | *** 1.21 nA |
17 | 3 | Jim Pulokas | ** 120 keV |
18 | *** 2.53 nA |
||
19 | 2 | Jim Pulokas | |
20 | |||
21 | h1. Exposure Time to Screen Current measurment |
||
22 | |||
23 | * refer to Tecnai Help for "Exposure Time, Plate Camera CP" |
||
24 | * formula: beam current (nA) = 2.15 * emulsion / measured_exposure_time |
||
25 | ** exp time is between 3.1 and 3.5 depending on size of beam |
||
26 | ** emulsion = 1 |
||
27 | 1 | Jim Pulokas | ** beam current is between 0.61 and 0.69 nA |
28 | 3 | Jim Pulokas | |
29 | h1. Leginon screen scale factor measurement |
||
30 | |||
31 | * adjust scale factor until beam current is same as faraday cup |
||
32 | ** 200 keV |
||
33 | *** 0.725 |
||
34 | ** 120 keV |
||
35 | |||
36 | h1. Camera sensitivity |
||
37 | |||
38 | ** 200 keV |
||
39 | *** 45-50 count/e |
||
40 | ** 120 keV |
||
41 | 1 | Jim Pulokas | *** 107 count/e |
42 | 4 | Jim Pulokas | |
43 | |||
44 | |||
45 | 6 | Jim Pulokas | h1. 2011-02-02 calibration on Tecnai 2: |
46 | 5 | Jim Pulokas | |
47 | 4 | Jim Pulokas | * faraday measurement: 1.52 nA |
48 | * Tecnai main screen measurement (emulsion = 1): 2.15 / 2.6 = 0.83 nA |
||
49 | 6 | Jim Pulokas | * Leginon scale factor to match faraday current: 0.793 |
50 | 7 | Jim Pulokas | * Legion measures Tietz sensitivity: 68.7 counts/e- |
51 | 9 | John Crum | |
52 | 10 | John Crum | h1. 2013-02-28 Calibration on T12 Spirit |
53 | |||
54 | * 120kV, Filament Heat= 29, step 1, Spot 3, mag 46K, Beam Current = 5.5-7.5 µA, LaB6 installed 2013-02-22 |
||
55 | * TVIPS F416 camera |
||
56 | * Technai main screen measurement (emulsion = 1): 2.15 / 0.11 = 19.54 nA |
||
57 | * Leginon scale factor to match Faraday current: 0.991 (was 0.80) |
||
58 | * Leginon measure Tietz sensitivity: 38.8294 counts/e- (was 32.5211) |
||
59 | |||
60 | 9 | John Crum | |
61 | h1. Finding the Faraday Cup |
||
62 | |||
63 | * See picture below |
||