h1. Faraday Cup Measurment Notes
2
3
Picoammeter setup:
4
* plug in
5
* wait 1 hour
6
* activate ZCHK (maybe already when turned on) (should say ZC on screen)
7
* plug into faraday holder
8
* set to lowest range (2nA)
9
* activate ZCOR to set zero  (should say ZZ on screen)
10
* set auto range if desired
11
* deactivate ZCHK to begin measuring
12 3 Jim Pulokas
* focus beam to a spot
13 11 Anchi Cheng
* pull out holder until spot hits the faraday cup (max current) (This is a negative value)
14
Examples
15 3 Jim Pulokas
** 200keV
16 11 Anchi Cheng
*** 1.21 nA
17 3 Jim Pulokas
** 120 keV
18
*** 2.53 nA
19 2 Jim Pulokas
20
21
h1. Exposure Time to Screen Current measurment
22
23
* refer to Tecnai Help for "Exposure Time, Plate Camera CP"
24
* formula:  beam current (nA) = 2.15 * emulsion / measured_exposure_time
25
** exp time is between 3.1 and 3.5 depending on size of beam
26
** emulsion = 1
27 1 Jim Pulokas
** beam current is between 0.61 and 0.69 nA
28 3 Jim Pulokas
29
h1. Leginon screen scale factor measurement
30
31
* adjust scale factor until beam current is same as faraday cup
32
** 200 keV
33
*** 0.725
34
** 120 keV
35
36
h1. Camera sensitivity
37
38
** 200 keV
39
*** 45-50 count/e
40
** 120 keV
41 1 Jim Pulokas
*** 107 count/e
42 4 Jim Pulokas
43
44
45 6 Jim Pulokas
h1. 2011-02-02 calibration on Tecnai 2:
46 5 Jim Pulokas
47 4 Jim Pulokas
* faraday measurement:  1.52 nA
48
* Tecnai main screen measurement (emulsion = 1):  2.15 / 2.6 = 0.83 nA
49 6 Jim Pulokas
* Leginon scale factor to match faraday current:  0.793
50 7 Jim Pulokas
* Legion measures Tietz sensitivity:  68.7 counts/e-
51 9 John Crum
52 10 John Crum
h1.  2013-02-28  Calibration on T12 Spirit
53
54
*  120kV, Filament Heat= 29, step 1, Spot 3, mag 46K, Beam Current = 5.5-7.5 µA, LaB6 installed 2013-02-22
55
*  TVIPS F416 camera
56
*  Technai main screen measurement (emulsion = 1): 2.15 / 0.11 = 19.54 nA
57
*  Leginon scale factor to match Faraday current:  0.991  (was 0.80)
58
*  Leginon measure Tietz sensitivity: 38.8294  counts/e- (was 32.5211)
59
60 9 John Crum
61
h1. Finding the Faraday Cup
62
63
* See picture below 
64
* Center to cup is 4mm
65
* Use a 6mm or 7mm Allen (hex) wrench as a wedge (spacer) between the holder end cap and stage.  Use the wrench as a lever to fine tune the position