Faraday Cup Measurment Notes » History » Version 8
John Crum, 08/27/2012 04:29 PM
1 | 1 | Jim Pulokas | h1. Faraday Cup Measurment Notes |
---|---|---|---|
2 | |||
3 | Picoammeter setup: |
||
4 | * plug in |
||
5 | * wait 1 hour |
||
6 | * activate ZCHK (maybe already when turned on) (should say ZC on screen) |
||
7 | * plug into faraday holder |
||
8 | * set to lowest range (2nA) |
||
9 | * activate ZCOR to set zero (should say ZZ on screen) |
||
10 | * set auto range if desired |
||
11 | * deactivate ZCHK to begin measuring |
||
12 | 3 | Jim Pulokas | * focus beam to a spot |
13 | * pull out holder until spot hits the faraday cup (max current) |
||
14 | ** 200keV |
||
15 | *** 1.01 nA |
||
16 | ** 120 keV |
||
17 | *** 2.53 nA |
||
18 | 2 | Jim Pulokas | |
19 | |||
20 | h1. Exposure Time to Screen Current measurment |
||
21 | |||
22 | * refer to Tecnai Help for "Exposure Time, Plate Camera CP" |
||
23 | * formula: beam current (nA) = 2.15 * emulsion / measured_exposure_time |
||
24 | ** exp time is between 3.1 and 3.5 depending on size of beam |
||
25 | ** emulsion = 1 |
||
26 | 1 | Jim Pulokas | ** beam current is between 0.61 and 0.69 nA |
27 | 3 | Jim Pulokas | |
28 | h1. Leginon screen scale factor measurement |
||
29 | |||
30 | * adjust scale factor until beam current is same as faraday cup |
||
31 | ** 200 keV |
||
32 | *** 0.725 |
||
33 | ** 120 keV |
||
34 | |||
35 | h1. Camera sensitivity |
||
36 | |||
37 | ** 200 keV |
||
38 | *** 45-50 count/e |
||
39 | ** 120 keV |
||
40 | 1 | Jim Pulokas | *** 107 count/e |
41 | 4 | Jim Pulokas | |
42 | |||
43 | |||
44 | 6 | Jim Pulokas | h1. 2011-02-02 calibration on Tecnai 2: |
45 | 5 | Jim Pulokas | |
46 | 4 | Jim Pulokas | * faraday measurement: 1.52 nA |
47 | * Tecnai main screen measurement (emulsion = 1): 2.15 / 2.6 = 0.83 nA |
||
48 | 6 | Jim Pulokas | * Leginon scale factor to match faraday current: 0.793 |
49 | 7 | Jim Pulokas | * Legion measures Tietz sensitivity: 68.7 counts/e- |