h1. Faraday Cup Measurment Notes
2
3
Picoammeter setup:
4
* plug in
5
* wait 1 hour
6
* activate ZCHK (maybe already when turned on) (should say ZC on screen)
7
* plug into faraday holder
8
* set to lowest range (2nA)
9
* activate ZCOR to set zero  (should say ZZ on screen)
10
* set auto range if desired
11
* deactivate ZCHK to begin measuring
12 3 Jim Pulokas
* focus beam to a spot
13
* pull out holder until spot hits the faraday cup (max current)
14
** 200keV
15
*** 1.01 nA
16
** 120 keV
17
*** 2.53 nA
18 2 Jim Pulokas
19
20
h1. Exposure Time to Screen Current measurment
21
22
* refer to Tecnai Help for "Exposure Time, Plate Camera CP"
23
* formula:  beam current (nA) = 2.15 * emulsion / measured_exposure_time
24
** exp time is between 3.1 and 3.5 depending on size of beam
25
** emulsion = 1
26 1 Jim Pulokas
** beam current is between 0.61 and 0.69 nA
27 3 Jim Pulokas
28
h1. Leginon screen scale factor measurement
29
30
* adjust scale factor until beam current is same as faraday cup
31
** 200 keV
32
*** 0.725
33
** 120 keV
34
35
h1. Camera sensitivity
36
37
** 200 keV
38
*** 45-50 count/e
39
** 120 keV
40 1 Jim Pulokas
*** 107 count/e
41 4 Jim Pulokas
42
43
44 6 Jim Pulokas
h1. 2011-02-02 calibration on Tecnai 2:
45 5 Jim Pulokas
46 4 Jim Pulokas
* faraday measurement:  1.52 nA
47
* Tecnai main screen measurement (emulsion = 1):  2.15 / 2.6 = 0.83 nA
48 6 Jim Pulokas
* Leginon scale factor to match faraday current:  0.793
49 7 Jim Pulokas
* Legion measures Tietz sensitivity:  68.7 counts/e-
50 9 John Crum
51
52
h1. Finding the Faraday Cup
53
54
* See picture below 
55
* Center to cup is 4mm
56
* Use a 6mm or 7mm Allen (hex) wrench as a wedge (spacer) between the holder end cap and stage.  Use the wrench as a lever to fine tune the position