cs is the coefficient of spherical aberration in meters. Find out from TEM manufacture the proper value for your tem.

zplane is a number that gives the order of camera. If a camera with low zplane value is going to acquire an image, Leginon will attempt to retract all camera at higher zplane values.

  • One computer for both tem and digital camera:
    [tem]
    class: tecnai.Tecnai
    cs: 2.0e-3
    [camera]
    class: gatan.Gatan
    zplane: 50
    
  • Sperate computers for the two instruments: configure only the instrument reside on the particular computer
    • On the computer that controls the tem:
      [tem]
      class: tecnai.Tecnai
      cs: 2.0e-3
      
    • On the computer that controls the digital camera:
      [camera]
      class: gatan.Gatan
      zplane: 50
      
  • Two camera controlled by the same computer and server such as in the case of DE12 and DE Survey camera that retract/insert at the same time.
    • On the computer that controls the digital camera:
      [camera1]
      class: de12.DE12
      zplane: 50
      [camera2]
      class: de12.DESurvey
      zplane: 50
      
  • Two camera is controlled by the same computer and server but one is retractable and one is non-retractable and at lower position.
    • On the computer that controls the digital camera:
      [camera1]
      class: de12.DE12
      zplane: 50
      [camera2]
      class: tietz.TietzPXL
      zplane: 5
      